Basic Specifications
| Display Block | 10.4-inch TFT color LCD Effective screen area: 211.2 mm x 158.4 mm (640 dots x 480 dots) | | Printer Block (RA1200/RA1300) | | | Printing method | Thermal printing using a thermal head | | | Paper Width | 219.5 mm | | | Effective Recording Width | 1 split (200mmFS) to 16 splits (10mmFS), Numberofsplit and printing width can be changed. | | | Channel Discrimination | Prints channel number in the vicinity of theprintedwaveform The ON/OFF function is available. | | Battery Backup | Backup Contents: Main unitsettings,Recording conditions Backup time: About 1 month (if fully charged and used undernormaltemperature) | | Storage Devices | A built-in 3.5-inch floppy diskdrive(1.25MB/1.44MB) PC card slot (ATA flash memory card, SRAM card) Built-in MO unit (640, 540, 230, and 128MB supported): optional | | Interface | RS-232C unit (38400 to 2400 bps): optional GP-IB unit (IEEE488 compliant): optional SCSI unit (50-pin type): optional | | Remote Terminal | Start/stop, mark print, and paper feed areavailableby external signal, synchronous pulse Input, protect input, output ofwaveformjudgement results, and error output | | Clock Function | Stability: Daily error within ±35 PPM (atnormaltemperature) | | Operating Environment | Temperature: 0 to 40°C (5 to 40°CwhenFFD is operating) Humidity: 35 to 80% RH (without condensation) Vibration Resistance: MIL-STD-810E compliant (except when FDorMOis operating) | | Power Supply | 90 to 132 VAC/180 to 264 VAC (to be specifiedwhenordering) Frequency 47 to 63Hz DC power unit (optional) In use: 11 to 28 VDC(AC powerisprioritized) | | PowerConsumption(Maximum) | RA1100: Approx. 85 VA RA1200: Approx. 180 VA (About 85 VA during stand-by) RA1300: Approx. 350 VA (About 85 VA during stand-by) | | Dimensions | Approx. 369.5 (W) x 150.5 (H) x 301(D)mm | | Weight | Approx. 6.6kg (RA1100 main unit withfour2-ch high-resolution DC amps) Approx. 7.5kg (RA1200 main unit with four 2-ch high-resolutionDCamps) Approx. 7.5 kg (RA1300 main unit with four 2-ch high-resolutionDCamps) |
Trigger Specifications | Trigger Mode | OR, AND, A*B, WINDOW | | Trigger Source | Input Signal, time trigger(startingtime, ending time, measurement interval), manual trigger and external trigger | | Trigger Setting | Amps other than event amps | | Trigger Slope | OR, AND.... .or | | | | A*B... or for each A,B source | | | | WINDOW...OUT or IN | Level Setting: To besetwith physical values (e.g. voltage) | Event Amp | | State Settings: | H, L, or OFF can be set for each input(withouttrigger slope) When OFF is set, trigger condition is notapplied. | | | Trigger Setting: | AND or OR of the statesettingconditions of inputs from 1 to 8 |
| | Trigger Related Functions | Pre-trigger , trigger pass count,triggerfilter, and trigger mark |
Other Specifications | PrintFunctions(RA1200/RA1300) | | | Data Information | Measurement mode,year/month/day,Measurement start time, Data No., Trigger conditions (Trigger point, Trigger year/month/day,Triggertime), Sampling speed, Chart speed, Time axis, etc. areprintedsimultaneously with waveform print). ON/OFF function is available. | | Channel Information | Input unit settings areprintedsimultaneously. ON/OFF function is available. | | Mark print | In the real-time mode, the mark printisavailable (M date/time) | | List Print. | Date, time, measurement mode, data No.,record form, chart speed, sampling speed, scale printing, back-light auto-off, autostart,buzzer click, trigger mode, source channel, recording operation,filingsetting printout | | Screen Copy Function | Prints hard copies of the contentsshownon the screen | | A4 report output | Registers free comments and printsthemtogether with the waveform printing | | Printing Line Width | Sets baseline boldness (1, 2, 3, 4dots)for each channel | | Key Lock Function | Invalidates key input topreventerroneous operation | | Storing and ReadingofSettings | Up to four conditions for input unitandmain unit settings are stored in the main unit memory | | PhysicalValueConversion | Physical conversion of inputsignal,full-scale modification of display | | Stand-by Function | When the main unit is restoredfromblackout or temporary halt, original state is regained.(Except when filing) | | AutomaticSetupFunction | By presetting a floppy diskcontainingthe measurement conditions in FD drive, the conditions are automatically read when turning onthepower. | | Storing theScreenImage | Stores a screen image in bitmapformatinto a floppy disk | | Help Function | Displays the handling method,cautions,etc. | | RS-232C Unit Function | | | Remote Control | Controls the main unit from a personal computerviathe RS-232C cable. | | FAX Transmission Function | Sends measured waveform by using a FAXmodemconnected to RS-232C. | | GB-IB Unit Function | | | Remote Control | Controls the main unit from a personal computerviaGB-IB cable. | | SCSI Unit Function | | | Stores data to MO or PD as externalmemorydevices. |
AC Bridge Power Supply Unit (Optional) | Function | Bridge power source for 2-ch ACstrainamp | | Power Voltage, Carrier Wave | 2 Vms, Sine wave 5kHz | | Synchronism | Synchronization with other RA1000susinga built-in AC bridge power unit is available. |
Memory Mode | Measurement Operation | Once, Repeat, Endless | | Memory Capacity | 256kW/CH (Max.1 MW/CHoptionallypossible) *When function operation is used, memory capacity is 1/4 When two channels are used, 2MW/CH is the maximum | | Memory Division | 1, 2, 4, 8, 16, 32, 64, and128divisions | | Sampling | 1 µs (1 MS/s) to 1 s(1S/s),User-settableExternal synchronization enabled | | Display(ReplayMonitor) | | | Waveform Display | Display magnification :x5 to x1/1000 | | X-Y Display | X-Y display where data of an anychannelis shown on the X axis, and other channel on the Y axis. Display step: 1 to 1000 steps | | Print (RA1200/RA1300) | | | Waveform Print | Print Density: Voltage axis: 8dots/mmTime axis: 10 dots/mm Copy magnification: x5 to x1/1000 | | Alphanumeric Print | Readout Interval : 1 to 1000 steps | | X-Y Print | X–Y print where data of an any channelisshown on the X axis, and other channel on the Y axis. Effective print range : 200 mm x 200mmPrint density : 400 dots (X axis) x 400 dots (Y axis) Display step : 1to1000 steps | | Memory Filing | Each time data is stored into memory,thedata is saved in the memory device in the binary or CSV format. |
Real-Time Mode (RA1200/RA1300) | Waveform Printing | | | Measurement Starting Operation | Starts with the Start key, by thetriggerdetection, or at set time. Interval recording available. | | Paper Speed | 25 mm/s to 1 mm/sec (RA1200), 100 mm/sto1 mm/min (RA1300), User settable. External synchronization enabled | | Frequency Response | DC to 100 kHz (sampling points:10points/cycle), Varies depending on the input units | | Printing Density | Voltage axis : 8 dots/mm | Time axis | :RA1200 | :10 dots/mm | | | :RA1300 | :20 dots/mm at 100 mm/s | | | | :40 dots/mm at 50 mm/s | | | | :80 dots/mm at 25 mm/s |
| | Printing Length | Continuous or 1 to 1000division-shotfeeding | | Alphanumeric Printing | | | Sampling | 1 s to 1 hour, User settable,Externalsynchronization enabled | | Printing Length Setting | Continuous or 1 to 1000 data shot feeding | | X-Y Printing | | | Performance | X-Y recording where data of ananychannel is shown on the X axis, and other channel on the Y axis. Continuous recording until stopbuttonis pressed (pose function available) | | Effective Printing Range | 200 mm x 200 mm | | Printing Density | 400 dots (X axis) x 400 dots (Y axis) | | Printing Speed | 10 ms or more | | Backup Filing | Stores the measured data in amemorydevice in real-time Recording Speed: Synchronized with paper speed |
Transient Mode | TransientRecording(RA1200/RA1300) | | | Function | Normally operates in the real-timemode.When trigger is detected, the mode is switched to the memory mode. | | Real-time Section | In conformance with the waveformprintingin the real-time mode | | Memory Section | In conformance with data storageandprinting in the memory mode | | Transient Filing | | | Function | Normally operates in backup filing,datais saved in a memory when trigger is detected, and whenmeasurementfinishes, the memory data is saved. | | Real-Time Section | In conformance with backup filing in thereal-timemode | | Memory Section | In conformance with memory filing inthememory mode |
Filing Mode | Data Recording | | | Function | Stores the measured data real-time toamemory device (Real time filing) | | Recording Speed | 200 µs to 60 s. Maximum recordingspeedvaries depending on number of channels and recording device. User settableExternalsynchronization ennobled | | Recording Method | Select normal or ring Ring recording is possible (Repeated recording during set time) | | WaveformPrinting(RA1200/RA1300) | | | Function | Prints the waveform of input signal | | Paper Speed | 25 mm/s to 1 mm/m (RA1200), 100 mm/s to1mm/m(RA1300) User Settable External synchronization enabled | | Frequency Response | DC to 100 kHz (sampling point:10points/cycle), Varies depending on the input units | | | Printing Density | Voltage axis : 8 dots/mm | Time axis | :RA1200 | :10 dots/mm | | | :RA1300 | :20 dots/mm at 100 mm/s | | | | :40 dots/mm at 50 mm/s | | | | :80 dots/mm at 25 mm/s |
|
FFT Unit (Optional) | 1 Signal Analysis | Linear spectrum, power spectrum, RMS spectrum,powerspectrum density, and octave analysis (1/1, 1/3) | | 2 Signals Analysis | Transfer function, coherencefunction,and cross-power spectrum | | Window Function | Rectangular, hanning,hamming,exponential, force | | Addition Processing | Simple additional average (timeaxis),simple additional average (frequency axis), exponential weighted average (frequencyaxis),and peak hold | | Analysis Frequency | 400kHz, 200kHz, 80kHz, 40kHz,20kHz,10kHz, 5kHz, 2kHz, 1kHz, 500Hz, 200Hz, 100Hz, 50Hz, 20Hz, and 10Hz | | Resolution | 400 (1000), 800 (2000), 1600 (4000) Numbers in parentheses indicates the number ofsamplingpoints. |
Arithmetic Operation Unit (Optional) | SectionalStatisticalOperation | Maximum value, minimum value, P-P value,averagevalue, area, effective value, standard deviation, risingtime,falling time | | Function Operation | Four-function calculation,absolutevalue, differential, integration, secondary differential,doubleintegration, square root, exponential, common logarithm,movingaverage, and trigonometric functions (sin, cos,tan, asin,acos,atan) |
Waveform Judgment Unit (Optional) | Function | Measures waveform of input signals ofeightchannels, and outputs the results of waveform judgement | | Targets ofOutputtingResults | Print on paper, FAX, data storage,orremote terminal output |

2-CH High Resolution Amp (AP11-101) | Input | 2channels/unit, isolated unbalanced input,isolatedBNC connector | | Input coupling | AC and DC coupling | | Input impedance | 1M or higher | | Measurement range | 0.1, 0.2, 0.5, 1, 2, 5, 10, 20, 50,100,200, and 500 V | | Range accuracy | Within 0.3% FS (Within 0.8% FS at 500V) | | Offset accuracy | Within 0.3% FS (at 25°C) | | Linearity | Within 0.1% FS | | Allowableinputvoltage | Range of 10 V to 500 V: 500 V max.(ACor DC peak values) Range of 0.1 V to 5 V: 100 V max. (AC or DC peak values) | | CMV | Unit only: 43 V (AC or DCpeakvalues)When using isolated BNC cable (optional): 300 VAC | | Frequency response | At DC coupling: DC to 50 kHz (+0.5,-3dB) At AC coupling: 0.3 to 50 kHz (+0.5, -3 dB) | | Low-pass filter | Bessel type (attenuation factor:-12dB/OCT) 30, 300, 3 kHz, and OFF (+0.5, -3 dB) | | A/D converter | 16 bits, 100 kHz max. (twochannelsimultaneous sampling) | | Temperature stability | Zero point: Within 0.03% FS/°C Gain (range): Within 0.01% FS/°C | | Weight | 230 g |
2-CH High Speed Amp (AP11-103) | Input | 2channels/unit, isolated unbalanced input,isolatedBNC connector | | Input coupling | AC and DC coupling | | Input impedance | 1M or higher | | Measurement range | 0.1, 0.2, 0.5, 1, 2, 5, 10, 20, 50,100,200, and 500 V | | Range accuracy | Within 0.5% FS (Within 0.8% FS at 500V) | | Offset accuracy | Within 0.5% FS (at 25°C) | | Linearity | Within 0.2% FS | | Allowableinputvoltage | Range of 10 V to 500 V: 500 V max.(ACor DC peak values) Range of 0.1 V to 5 V: 100 V max. (AC or DC peak values) | | CMV | Unit only: 43 V (AC or DCpeakvalues)When using isolated BNC cable (optional): 300 VAC | | Frequency response | At DC coupling: DC to 400 kHz (+0.5,-3dB) At AC coupling: 0.3 to 400 kHz (+0.5, -3 dB) | | Low-pass filter | Bessel type (attenuation factor:-12dB/OCT) 50, 500, 5 kHz, and OFF (+0.5, -3 dB) | | A/D converter | 12 bits, 1 MHz max. (twochannelsimultaneous sampling) | | Temperature stability | Zero point: Within 0.03% FS/°C Gain (range): Within 0.01% FS/°C | | Weight | 240 g |
2-CH TC/DC Amp (AP11-106) | Input | 2channels/unit, isolated unbalanced input,terminalblock 4 | | Input coupling | DC coupling | | Input impedance | 10M or higher | | Thermocouple | R, T, J, K, and W | | Measurement range | R: 0 to 1600°C T: -200 to 400°C J: -200 to 1000°C K: -200 to 1350°C W: 0 to 2300°C | | Voltagemeasurementrange | 100, 200, 500 mV, 1, 2, 5, 10, 20 and50V FS | | Range accuracy | Temperature: Within 0.5% FS (Within1%FS at 0°C or lower) Voltage: Within 0.5% FS | | Coldjunctioncompensation | Internal/external switchable Accuracy: Within 2°C (Within 1°C at stable temperature of 20°Catinput terminal) | | Linearity | Within 0.1% FS | | Allowableinputvoltage | 50 V (AC or DC peak values) | | CMV | 43 V (AC or DC peak values) | | Frequency response | DC to 40 kHz (+0.5, -3 dB) | | Low-pass filter | Bessel type (attenuation factor:-18dB/OCT) 1, 30, 500, 5 kHz, and OFF (+0.5, -3 dB) | | A/D converter | 15 bits, 100 kHz max. (twochannelsimultaneous sampling) | | Temperature stability | Accuracy: 0.04% FS/°C (When usedastemperature amp) Zero point: Within 0.03% FS/°C (When used as DC amp) Gain (range): Within 0.01% FS/°C | | Weight | 240 g |
TC/DC Amp (AP11-107) | Input | 1 channel/unit, isolated unbalanced input, 2bindingposts | | Input coupling | DC coupling | | Input impedance | 10M or higher (1M at 5, 10, 20, 50 V FS in DC amp) | | Thermocouple | R, T, J, K | | Measurement range | R: 0 to 800°C at 800°C FS and 0to1600°C at 1600°C FS T: -200 to 200°C at 200°C FS and -200 to 400°C at 400°CFS J: -200 to 200°C at 200°C FS and -200 to 1000°C at 1000°C K: -200 to 200°C at 200°C FS and -200 to 1200°C at 1200°C | | Voltagemeasurementrange | 10, 20, 50, 100, 200, 500 mV, 1, 2,5,10, 20, and 50 V FS | | Range accuracy | Temperature: Within 0.5% FS (Within1%FS at 0°C or lower) Voltage: Within 0.5% FS | | Linearity | Within 0.1% FS | | Allowableinputvoltage | 50 V (AC or DC peak values) | | CMV | 300V (AC or DC peak values) | | Frequency response | DC to 40 kHz (+0.5, -3 dB) | | Low-pass filter | Bessel type (attenuation factor:-18dB/OCT) 1, 30, 500, 5 kHz, and OFF (+0.5, -3 dB) | | Referencejunctioncompensation | Internal/external switchable | | A/D converter | 14 bits, 100 kHz max. | | Weight | 200 g |
Event Amp (AP11-105) | Number of channels | 8 channels/unit | | Input signal | Sets voltage or contact foreachchannel Voltage input: Input voltage range 0 to +24 V | Detection level | H level...2.5 V or higher | | | L level...0.5 V or lower | | Contact input: | Close...250 or lower | | | Open...2 k or higher |
| | Response time | 1µs or faster (Input level should be+5V or higher.) | | Input connector | Round DIN connector 8P...2 pcs. Event amp side: XT2B-0800(Conformed to DIN45326) | | Weight | 100 g |
2-CH AC Strain Amp (AP11-104) | Number of channels | 2 channels/unit, isolated balanced input,NDISconnector | | Applicablegaugeresistance | 120 to 1k | | Bridge power supply | Sine wave 2 Vrms, 5 kHz (Bridgepowersupply is supplied separately.) | | Gauge factor | 1.9 to 2.2 | | Range of balance | Resistance: 2% (10000 µ )or lower Capacitance: 2000 pF or lower | | Balance method | Resistance: Auto-balance Capacitance: Auto-balance (500 pF or lower,eliminatingconstantly) Balance time: Within 1s at 1 ch Within 0.5%Fs Remainedvoltageaccuracy | | Maximum sensitivity | Over full-scale at 500 µ (at 2 V or higher bridge voltage) | | Measurement range | 1, 2, 5, 10, 20 kµ FS | | Internalcalibratorand accuracy | 0.5 k, 1 k, 2 k, 3 k, 5 kµ Accuracy: Within 0.5% FS | | Linearity | 0.2% FS | | CMV | 300 VAC | | Frequency response | DC to 2 kHz (+0.5, -3dB) | | Low-pass filter | Butterworth type (attenuationfactor:-12 dB/OCT) 10, 30 , 100, 300 Hz, and OFF (+0.5, -3 dB) | | A/D converter | 16 bits, 100 kHz max. | | Temperature stability | Zero point: Within 0.05% FS/°C Gain (range): Within 0.05% FS/°C | | Weight | 285 g |
2-CH DC Strain Amp (AP11-110)
| Number of channels | 2 channels/unit, isolatedbalancedinput, NDIS connector | | Input coupling | DC | | Input impedance | 10M +10M or higher | | Bridge power supply | 2 V and 5 V | | Applicablegaugeresistance | 120 to 2 k (at BV = 2V),350 to 2k (at BV=5 V) | | Gauge factor | 2 | | Range of balance | 2% (15000 µ orlower | | Balance method | Resistance: Auto-balance Balance time: 0.5 sec max. at 1ch Remained voltage accuracy: Within 0.3% FS | | Measurement range | 2, 5, 10, 20, 50 kµ FS (at BV = 2 V) 0.8, 2, 4, 8, 20 kµ FS(BV= 5 V) | | Voltagemeasurementrange | 2, 5, 10, 20, 50 mV FS | | Range accuracy | Within 0.3% FS | | Linearity | 0.1% FS | | Allowableinputvoltage | 8 V (DC or AC peak values) | | Frequency response | DC to 50 kHz (+0.5, -3dB) | | Low-pass filter | Bessel type (attenuation factor:-12dB/OCT) 10, 30, 300 Hz, 1 kHz, and OFF (+0.5, -3 dB) | | Temperature stability | Zero point: Within 0.1% FS/°C Gain (range): Within 0.01% FS/°C | | CMV | 300 VAC | | A/D converter | 16 bits, 100 kHz max | | Weight | 240 g |
2-CH Vibration/RMS Amp (AP11-109) | Input | 2 channels/unit, isolatedunbalancedinput, isolated BNC connector | | Input coupling | AC and DC | | Input impedance | 10 M orhigher | | Power supplyforsensor | 2 mA, 18 V or higher | | Measurement range | 0.1, 0.2, 0.5, 1, 2, 5, 10, 20, 50,100,200, 500 V | | Range accuracy | DC amp range: Within 0.3% FS(Within0.8% FS at 500 V) RMS amp range: Within 2% FS (at DC and 40 Hz to 20 kHz) | | Linearity | Within 0.1% FS | | Crest factor | 2.8 max. (RMS amp) | | CMV | Unit only: 42 V (DC or ACpeakvalues) When isolated BNC cable (optional) is used: 300 V AC | | Frequency response | DC coupling: DC to 50 kHz (+1,-3dB) AC coupling: 1 to 50 kHz (+1, -3 dB) | | Low-pass filter | Butterworth type (attenuationfactor:-24 dB/OCT) 30, 100, 300 Hz, 1 kHz, and OFF | | High-pass filter | Butterworth type (attenuationfactor:-24 dB/OCT) 10, 30, 100 Hz, and OFF | | A/D converter | 16 bits, 100 kHz max. | | Temperature stability | Zero point: Within 0.02% FS/°C Gain (range): Within 0.01% FS/°C | | Weight | 270 g |
2-CH FFT Amp (AP11-102) | Input | 2 channels/unit, isolatedunbalancedinput, isolated BNC connector | | Input coupling | AC and DC (only AC couplingwhenamp-embedded charge accelerometer is used) | | Input impedance | 1 M or higher | | Power supplyforsensor | 2 mA, 18 V or higher | | Measurement range | 0.1, 0.2, 0.5, 1, 2, 5, 10, 20, 50,100,200, 500 V | | Range accuracy | Within 0.3% FS (Within 0.8% FS at 500V) | | Linearity | Within 0.1% FS | | Maximum input voltage | 500 V (DC or AC peak values) (30 V atACcoupling of 0.1 to 5 V) | | CMV | Unit only: 42 V (DC or ACpeakvalues) When isolated BNC cable (optional) is used: 300 V AC | | Frequency response | DC coupling: DC to 50 kHz (+0.5,-3dB) AC coupling: 0.3 to 50 kHz (+0.5, -3 dB) | | Low-pass filter | Bessel type (attenuation factor:-24dB/OCT) 30, 300 Hz, 3 kHz, and OFF (+0.5, -3 dB) | | Anti-aliasing filter | 20, 40, 80, 200, 400, 800 Hz, 2, 4,8,20, and 40 kHz Drop characteristic: -72 dB typ. at 1.5 x fc | | Offset accuracy | Within 0.3% FS (at 25 °C) | | A/D converter | 16 bits, 100 kHz max. | | Temperature stability | Zero point: Within 0.02% FS/°C Gain (range): Within 0.01% FS/°C | | Weight | 240 g |
F/V Converter (AP11-108) | Input | 1 channel/unit, isolatedunbalancedinput, isolated BNC connector | | Input coupling | AC and DC | | Input impedance | 100 M or higher | | Input frequency range | 1 Hz to 10 kHz (Pulse width: 20 µsorlonger) | | Measurement range | 0.1, 0.2, 0.5, 1, 2, 5, 10 kHz FS | | Range accuracy | Within 0.5% FS (Within 0.8% FS at 500V) | | Linearity | Within 0.3% FS | | Trigger level | Selectable from 0 V and 2.5 V | | Allowableinputvoltage | 100 V (AC or DC peak values) | | CMV | Unit only: 42 V (AC or DCpeakvalues) Between isolated BNC and safety terminal (optional): 300 VAC | | Response time | Approx. 20 ms (at the range of 10 kHz) | | A/D converter | 16 bits, 100 kHz max. | | Temperature stability | Zero point: Within 0.03% FS/°C Gain (range): Within 0.02% FS/°C | | Weight | 240 g |
Charge Converter (AP11-901)
| Gain | 1.0 mV/pC ±5% | | Max. input charge | 5000 pC | | Frequency range | Approx. 1.6 Hz to 50 kHz | | Max. output voltage | 5 Vp-p or lower | | Drive voltage | 12 to 25 VDC | | Drive current | 0.5 to 5.0 mA | | Rated noise | 20 µVrms or lower | | Phase | 180°(Inverted input is output.) | | Operating temperature | -20 to 80°C | | Connector | Input: Miniature connector (10-32) Output: Male BNC terminal | | Weight | 20 g |
Charge Converter (AP11-902) | Gain | 1.0 mV/pC ±5% | | Max. input charge | 5000 pC | | Frequency range | Approx. 1.6 Hz to 50 kHz | | Max. output voltage | 5 Vp-p or lower | | Drive voltage | 12 to 25 VDC | | Drive current | 0.5 to 5.0 mA | | Rated noise | 20 µVrms or lower | | Phase | 180°(Inverted input is output.) | | Operating temperature | -20 to 80°C | | Connector | Input: Miniature connector (10-32) Output: Male BNC terminal | | Weight | 65 g |
Charge Converter (AP11-903)
| Gain | 1.0 mV/pC ±5% | | Max. input charge | 50000 pC | | Frequency range | Approx. 1.6 Hz to 50 kHz | | Max. output voltage | 5 Vp-p or lower | | Drive voltage | 12 to 25 VDC | | Drive current | 0.5 to 5.0 mA | | Rated noise | 20 µVrms or lower | | Phase | 180°(Inverted input is output.) | | Operating temperature | -20 to 80°C | | Connector | Input: Miniature connector (10-32) Output: Male BNC terminal | | Weight | 65 g |

| | Name | Model | Rating | | Main unit | Omniace II | RA1100 | Data acquisition system without printer | | RA1200 | Data acquisition system with printer, paperspeed:25 mm/s max. | | RA1300 | Data acquisition system with printer, paperspeed:100 mm/s max. |
| Standard accessories | AC power cable x 1, Recording paper x1(RA1200/1300), Paper holder x 1 (RA1200/1300), Slot cover plateforinput and interface units x 1 set, and user's manual x1set |
|
|